1 | Select the primary calibration data set.
|
2 | Select Edit. This opens the Beam Mode dialog box for the appropriate beam mode. In edit mode you will be able to edit, create and delete beam modes.
|
3 | Open the Calibration page. The Calibration page displays all calibration data sets made and a short description of them.
|
4 | Select Merge calibration for 2 spheres. The Calibration Merge dialog box opens.
Note
In cases where both calibrations contain gain adjust for the same beam, the value from the primary calibration will be used
in the resulting, merged calibration data set.
|
5 | Enter the name of the merged calibration data. |
6 | Select Merge selected calibrations. |
7 | Select the OK or Apply. The resulting calibration data set will be stored and the coefficients in the formulas used for processing data will be adjusted.
|