1 | Start the ME70 program in “inactive”mode. The mode of the ME70 program is visible in the Setup menu.
|
2 | On the Setup menu, select Calibration.![]() |
3 | Select File in the Calibration menu system. Select the replay XML file from the chosen calibration you want to edit or examine.
|
4 | Open the Setup menu. Check range limits and sphere
|
5 | Select Start in the Calibration menu system. This will load the calibration data into the calibration system.
|
6 | Select Stop in the Calibration menu system. This is done when the calibration data has been loaded.
|
7 | Open the TS Data tab in the Numerical information field. Suspend invalid data for each beam in this view.
|
8 | Select Save As in the Calibration menu. Generate new files. This stores the edited data. Load the coefficients into the mode.
|